Medtronic CareLink 2090 Programmer

FDA Recall Z-0406-2019 · Posted None

Recall StatusOpen, Classified
Product CodeOSR
Date Initiated2018-10-02
Product Quantity30172 units

Reason for Recall

Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download.

Firm Action

The firm began disseminating the Field Corrective Action notices by hand delivery on 10/02/2018 and followed with the posting of Medtronic Security Bulletin and the Urgent Medical Device Correction customer letter online on 10/11/2018. The notice explained vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download and beginning October 11th, 2018, Medtronic will be disabling the SDN for programmer updates and will rely solely on the USB update method.

Root Cause

Software design

Distribution Pattern

Worldwide