Medtronic CareLink Encore 29901 Programmer
FDA Recall Z-0407-2019 · Posted None
Advertisement
Recalling FirmMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Recall StatusOpen, Classified
Product CodeOSR
Date Initiated2018-10-02
Product Quantity4072 units
Reason for Recall
Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download.
Firm Action
The firm began disseminating the Field Corrective Action notices by hand delivery on 10/02/2018 and followed with the posting of Medtronic Security Bulletin and the Urgent Medical Device Correction customer letter online on 10/11/2018.
The notice explained vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download and beginning October 11th, 2018, Medtronic will be disabling the SDN for programmer updates and will rely solely on the USB update method.
Root Cause
Software design
Distribution Pattern
Worldwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.