MEVION S250 Product Usage: Proton Radiation Therapy

FDA Recall Z-0409-2017 · Posted 2016-11-15

Recall StatusTerminated
Product CodeMUJ
Date Initiated2016-10-31
Date Terminated2017-06-29
Product Quantity6 units

Reason for Recall

Completing Setup Without Program Move Complete Can Cause An Isocenter Error

Firm Action

Mevion issued an Important Safety Notice letter on October 31, 2016 to customers. The identified the affected product, problem, mitigation and advisory information. The letter informed customers that Mevion will release and install a software patch that removes the reliance on tracking isocenter changes for the correct reporting of the Couch position to the user displays and to the OIS. For further information mailto:[email protected]

Root Cause

Software design

Distribution Pattern

US Nationwide in the states of FL, MO, NJ and OK