The Peripheral Cutting Balloon Device small Monorail Delivery System features a balloon with 3 or 4 atherotomes (microsurgical blades) mounted longitudinally on its outer surface. When the Peripheral Cutting Balloon device is inflated, the atherotomes score the plaque, creating initiation sites for crack propagation. Percutaneous Transluminal Angioplasty (PTA) with the Peripheral Cutting Balloon Device allows dilatation of the target lesion with less pressure.

FDA Recall Z-0410-06 · Posted 2006-02-01

Recall StatusTerminated
Product CodeLIT
Date Initiated2005-12-07
Date Terminated2007-08-10
Product Quantityindividual qty in US:376, Worldwide distribution pending.

Reason for Recall

The distal shaft of the catheter may separate during withdrawal of the device from the patient. If the shaft separates, the surgical procedure may be prolonged or require more in-depth surgery to physically remove the broken-off shaft from the blood vessel.

Firm Action

All affected direct accounts will be given notification by Overnight mail. A 100% of the total number of customer are to be contacted.

Root Cause

Other

Distribution Pattern

Nationwide plus Argentina Austria Belgium Brunei Canada Chile China Colombia Czech Republic Denmark Finland France Germany Great Britain Hong Kong Hungary India Indonesia Ireland Israel Italy Latvia Lebanon Luxembourg Macau Malaysia Mexico Netherlands Norway Poland Portugal Puerto Rico Saudi Arabia. Singapore South Africa Spain Sweden Switzerland & Turkey