MEVION S250 Product Usage: Proton Radiation Therapy System
FDA Recall Z-0411-2017 · Posted 2016-11-15
Advertisement
Recalling FirmMevion Medical Systems, Inc.
Recall StatusTerminated
Product CodeMUJ
Date Initiated2016-10-20
Date Terminated2017-05-15
Product Quantity6 units
Reason for Recall
Software defect:2D projection of contours in Verity. The defect causes structures to appear off from their true position in the DRR and the radiograph with an error that increases with the distance of the structure from isocenter
Firm Action
Mevion issued an Important Safety Notice letter on October 14, 2016 to their customers. The letter identified the affected product, problem, containment, mitigation and actions taken. Customers were informed that Mevion will be updating software when it is released. For further Information, contact: [email protected].
Root Cause
Software design
Distribution Pattern
US Nationwide Distribution in the states of: FL, MO, NJ, OH, OK