SSCOR Suction Device Model 2310V
FDA Recall Z-0412-05 · Posted 2005-01-27
Advertisement
Recalling FirmS S C O R Inc
Recall StatusTerminated
Product CodeBTA
Date Initiated2005-01-06
Date Terminated2011-06-13
Product Quantity16
Reason for Recall
Reports from medical professionals that SSCOR suction device has not performed as intended. Firm's investigation disclosed that the condition in questions is caused by cracked regulators that result in inadequate suction.
Firm Action
SSCOR Powered Suction Device not performing as intended.
Root Cause
Other
Distribution Pattern
CA, FL, IA, IL, IN, KY, MO, NC, NJ, NY, OH, France, New Zealand, Singapore, United Kingdom
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.