Terminated

Spine Frontier Inspan compressors Part Number: 11-60004. Inspan compressors provide pressure to drive spikes of Inspan implants into the spinous process bone, resulting in fixation of the implant.

FDA Recall Z-0412-2013 · Posted 2012-11-21

Recalling FirmSpineFrontier, Inc.
Recall StatusTerminated
Product CodeKWP
Date Initiated2012-08-28
Date Terminated2015-06-11
Product Quantity17 units

Reason for Recall

Inspan Compressors may break at weld causing a loss of compression

Firm Action

Spine Frontier issued e-mail notices to customers on May 14, 2012 advising of the potential for breakage. Units were replaced with weld free design.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution-USA (nationwide) including PR and the states of FL, MA, MO, OK, and TX, and the country of Mexico.