Smith & Nephew ACUFEX Rap-Pac ACL Accessory Kit, Sterile Catalog Number: 014725
FDA Recall Z-0413-03 · Posted 2003-01-09
Advertisement
Recalling FirmSmith & Nephew, Inc.
Recall StatusTerminated
Product CodeHRX
Date Initiated2002-11-08
Date Terminated2005-04-20
Product Quantity77 units
Reason for Recall
Potential for non sterility as package tray does not contain aeration holes for ETO processing.
Firm Action
Smith & Nephew notified customers by letter on 11/8/02. Users are requested to remove stock and return to Smith & Nephew.
Root Cause
Process control
Distribution Pattern
Nationwide: AL, CA, IL, KY, FL, GA, IN, ME, MI, NH,NY, OH, SC, TX, WV,
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.