Merit Medical Systems, Inc., Convenience Trays, custom packaged under the following Merit Medical labels: (1) Prentice PICC Kit, Catalog number K12T-03698; (2) Insertion Kit, catalog number K12T-03050C; and (3) Angio Procedure Pack, catalog number K09T-09530A.

FDA Recall Z-0413-2010 · Posted 2009-11-23

Recall StatusTerminated
Product CodeKDD
Date Initiated2009-05-19
Date Terminated2009-12-03
Product Quantity378 units

Reason for Recall

Surgical gloves in hospital convenience kits may be degraded and tear during use.

Firm Action

Merit Medical Systems notified customers by phone, fax, or visit by sales representatives on 05/19/09. The notification advised of the convenience kit correction due to the recall of the Derma Prene Ultra Surgical Gloves, and that use of the product should cease immediately. Sales representatives were instructed to provide replacement gloves and "warning" stickers to be placed on identified kits in inventory. The warning stickers advise the customer to discard the surgical gloves in the kit and replace. Customers were requested to verify product disposition by signing a response form.

Root Cause

Nonconforming Material/Component

Distribution Pattern

USA distribution: nine hospitals in Texas and Colorado.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.