Elements Obturation Unit (EOU) used with respective Buchanan Heat Pluggers Device is an endodontic root canal plugger. It is intended to be used in Dentistry to provide continuous heat at the tip of a dental instrument to test tooth response to thermal stimulus, for tissue cauterization and coagulation, to backfill and downpack gutta percha during Endodontic root canal treatment, and when in Extruder mode it is used only to backfill gutta percha during root canal obturation.

FDA Recall Z-0413-2018 · Posted None

Recalling FirmOrmco/Sybronendo
Recall StatusTerminated
Product CodeEKR
Date Initiated2017-12-27
Date Terminated2024-04-29
Product Quantity25,765 total

Reason for Recall

Buchanan Heat Pluggers will possibly overheat while used with the Elements Obturation Unit (EOU).

Firm Action

Sybron Endo sent an Urgent: Medical Device Recall letter dated December 27, 2017. Customers were informed that Sybron Endo was still evaluating the root cause and would provide additional instructions once the root cause is determined. Distributors were asked to cease distributing any Elements Obturation Units (EOU), confirm if any are in inventory, return any affected devices, provide the notification to customers if further distributed, return any affected devices received from customers, and complete and return the acknowledgement form. Customers were instructed to examine current inventory for affected product, check the heat plugger to see if the NiTi icon appears and/or the tip turns glowing red, if it does do not use the heat plugger and return it to Sybron Endo, and complete and return the acknowledgement form. For further questions, please call the Customer Care department at (800) 537-7123.

Root Cause

Under Investigation by firm

Distribution Pattern

USA (nationwide) distributed to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Distributed internationally to Armenia, Australia, Canada, Chile, Ecuador, El Salvador, Guatemala, Hong Kong, India, Japan, Malaysia, Mexico, Netherlands, Pakistan, Panama, Philippines, Singapore, South Korea, Switzerland, Thailand, United Kingdom, Vietnam.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →