GE Millennium Systems: MG. Multi-Geometry Digital CSE Dual Detector Gamma Camera. A modular nuclear medicine imaging system. The MG gantry allows the detectors to be oriented in the 180-degree and 101.25-degree orientations. The other modular core components include a 3-axis universal imaging table, a GENIE Acquisition station, dual Digital CSE (Correlated Signal Enhancement) detectors and high-performance collimators. Features optional real-time body-contoured SPECT and whole body scans, as well as ACuscan, an optional transmission attenuation correction system for SPeC and Gated SPECT scans.

FDA Recall Z-0414-06 · Posted 2006-01-24

Recalling FirmGE Healthcare
Recall StatusTerminated
Product CodeKPS
Date Initiated2005-04-18
Date Terminated2008-02-24

Reason for Recall

In certain cases, one of the four collimator locking mechanisms that attach the collimator to the head may work itself free if collimator exchange has not been performed for an extended period of time. This situation, if neglected, could eventually cause the locking mechanism to unlock itself.

Firm Action

Customer letter is being sent to all customers who have an affected system. The letter alerts customers of the potential release of the locking mechanism and requires a Weekly Collimator Exchange Procedure to avoid release of the locking latch until the permanent solution is made available.

Root Cause

Other

Distribution Pattern

nationwide including Puerto Rico, and to countries including: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Denmark, Egypt, France, Germany, Iceland, India, Israel, Italy, Jordan, Korea, Kuwait, Luxembourg, Martinique, Mexico, New Zealand, Philippines Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Taiwan and United Kingdom.