AutoSuture Bladed Trocar, equipped with safety shield. Model number 179094. Available with smooth or threaded sleeve in sizes 5-15mm inner diameter and 70-100mm length.

FDA Recall Z-0414-2007 · Posted 2007-02-08

Recall StatusTerminated
Product CodeNLM
Date Initiated2006-10-19
Date Terminated2008-10-31
Product Quantity1566

Reason for Recall

The tip of the trocars may separate before and while in use. In some instances the trocars have been found with exposed blades upon opening the packaging.

Firm Action

The recall was initiated on 10/19/2006 by direct distribution of the recall notice to direct accounts by the firm's sales force. The firm also posted the notification on their web site. All customers were instructed to examine their inventories for the product and to remove and quarantine all affected product. Instructions on returning the product to the firm were also included along with a response form.

Root Cause

Process design

Distribution Pattern

Nationwide

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →