AIMER ENDOFEMORAL 6MM OFFSET (GREEN) Part Number 72201716. The product is an orthopedic device.

FDA Recall Z-0416-2010 · Posted 2009-11-24

Recall StatusTerminated
Product CodeNBH
Date Initiated2009-09-03
Date Terminated2012-05-08

Reason for Recall

Device may have improper Laser Marking, 4mm device may be marked as 6mm and 6mm device may be marked as 4mm.

Firm Action

A notification letter was issued via Fedex on September 3, 2009 to domestic consignees and international consignees were notified via e-mail. The letter described the issue as the affected device may have improper laser marking. The 4mm devices are marked as 6mm and the 6mm devices are marked at 4mm. Customers were instructed to return the completed response form and affected product in inventory to Smith & Nephew. Direct questions about the recall to your local Smith & Nephew Sales Representatives or call 1-508-261-3655.

Root Cause

Labeling design

Distribution Pattern

Worldwide Distribution -- United States including Puerto Rico, Australia, Austria, Belgium, Canada, Costa Rica, Dubai, Finland, France, Germany, Great Britain, Netherlands, Norway, Portugal, Spain, Sweden, and Switzerland.