Terminated

Access Immunoassay Systems Thyroglobulin Antibody Reagent Kit, Part Number 33890

FDA Recall Z-0417-2007 · Posted 2007-02-07

Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeDDC
Date Initiated2006-10-26
Date Terminated2012-04-30
Product Quantity474 units

Reason for Recall

Low end imprecision affecting both S0 calibrators and patient samples which can lead to inaccurate low level patient results when using the Access Thyroglobulin Antibody assay.

Firm Action

A Product Corrective Action (PCA) letter was mailed on Oct 27, 2006 to all Thyroglobulin Antibody customers. They are informed that there has been an increase in customer feedback regarding low end imprecision affecting both S0 calibrators and patient samples, which can lead to inaccurate low level patient results when using the Access Thyroglobulin Antibody assay. The letter informs the customer that the firm recommends that they use an alternate method for testing TgAb. (Letter was sent by US mail). Customers are asked to complete and return the enclosed response form within 10 days.

Root Cause

Other

Distribution Pattern

Nationwide and Canada