MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: MA3302-PM. CATALOG #(s): MA3302-PM
FDA Recall Z-0419-2018 · Posted None
Advertisement
Recalling FirmStryker Medical Division of Stryker Corporation
Recall StatusTerminated
Product CodeDWJ
Date Initiated2017-07-21
Date Terminated2018-03-22
Product Quantity11160
Reason for Recall
It was identified that the Mistral-Air Premium blankets may be capable of
conducting electricity during a defibrillation or cardioversion therapy at higher
energy levels (70 Joules or more) when the blanket is applied with the reflective
side toward the patient (upside down).
Firm Action
Stryker notified customers of the recall by sending Customer Notification and Business Reply Forms, dated July 21, 2017.
Root Cause
Component design/selection
Distribution Pattern
U.S.