LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 5.5mm x 55mm; f. 6.5mm x 35mm; g. 6.5mm x 40mm; h. 6.5mm x 45mm; i. 6.5mm x 50mm; j. 6.5mm x 55mm; k. 6.5mm x 60mm; l. 7.5mm x 35mm; m. 7.5mm x 40mm; n. 7.5mm x 45mm; o. 7.5mm x 50mm; p. 7.5mm x 55mm; q. 7.5mm x 60mm; r. 8.5mm x 35mm; s. 8.5mm x 40mm; t. 8.5mm x 45mm; u. 8.5mm x 50mm; v.8.5mm x 55mm; w. 8.5mm x 60mm; Used in spinal fixation surgery.

FDA Recall Z-0419-2023 · Posted 2022-12-05

Recall StatusTerminated
Product CodeNKB
Date Initiated2022-01-27
Date Terminated2023-10-23
Product Quantity1,702 screws

Reason for Recall

Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.

Firm Action

Integrity Implants notified consignees on 01/28/2022, via emailed letter. The letter instructed consignees to quarantine affected product, complete and return the Product Tracking & Verification Form, and return affected product. On 02/05/2022 the recall was expanded, and consignees were notified via emailed letter. The listed the additional affected product and contained the same instructions to consignees.

Root Cause

Device Design

Distribution Pattern

Nationwide distribution to AZ, CA, CT, DC, FL, IL, IN, LA, MA, MN, NJ, OK, PR, TN, TX, VA.