Terminated

Pharmaseal Thoracentesis Tray with Catheter, catalog number 4341B; a sterile, single use, Rx, latex free procedure tray containing a 7'' long Thoracentesis Catheter with a 14 gauge x 2'' needle, catheter guard, 60 ml syringe, needles, 5 ml syringe, drainage tube with needle, fluid collection bag, three prelabeled specimen vials with caps, 2 swab sticks, three gauze pads, fenestrated drape with adhesive strips, towel, 5 ml -1% lidocaine hydrochloride, hospital wrap and bandage; Allegiance Healthcare Corporation, McGaw Park, IL 60085 USA

FDA Recall Z-0420-03 · Posted 2003-01-09

Recalling FirmCardinal Health
Recall StatusTerminated
Product CodeGCB
Date Initiated2002-11-15
Date Terminated2006-04-14
Product Quantity232,228 trays

Reason for Recall

Raw material used in the manufacture of catheters may contain a contaminant predisposing the catheter to become brittle, possibly breaking during use.

Firm Action

Recall letters dated 11/15/02 were sent via 2-day FedEx to the direct accounts. The accounts were informed of the potential for breakage of the catheters during use, requested to inspect their inventories for the affected lot numbers, and segregate any found for return to Allegiance for replacement. Distributors were requested to sub-recall the product from their accounts. The recall was extended to include the 21 additional lots by letter dated 4/8/04. The letters were sent via Airborne Express, 1-Day, on 4/8/04. The instructions were the same as the 11/15/02 letter, except that Allegiance has changed their name to Cardinal Health.

Root Cause

Other

Distribution Pattern

Nationwide distribution, including Puerto Rico and Guam, and internationally to Saudi Arabia

Official Source

View FDA Recall →