Terminated

LineSider Spinal System pedicle screws, 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Model Number LS-N6SG6545. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Used in spinal fixation surgery.

FDA Recall Z-0421-2023 · Posted 2022-12-05

Recall StatusTerminated
Product CodeNKB
Date Initiated2022-08-03
Date Terminated2023-10-23
Product Quantity50 screws

Reason for Recall

Potential for intraoperative screw head separation at weld location.

Firm Action

Integrity Implants notified consignees on about 08/03/2022 via emailed letter. The letter instructed consignees to examine inventory and quarantine any affected product, arrange for the return of affected products, and complete and return the response form.

Root Cause

Under Investigation by firm

Distribution Pattern

US distribution to California, Indiana, New York, Oklahoma, Washington DC

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.