Trabecular Metal Tibial Impactor. Product Usage: TM Tibial Impactor pads are used in conjunction with the TM Tibial Impactor to impact the tibial tray onto the tibia.

FDA Recall Z-0423-2015 · Posted 2014-11-24

Recalling FirmZimmer, Inc.
Recall StatusTerminated
Product CodeHWA
Date Initiated2014-10-14
Date Terminated2015-06-09
Product Quantity1165 units

Reason for Recall

To provide clarifying instructions relating to the assembly/disassembly of the Trabecular Metal Tibial Impactor with/from TM tibial Impactor Pads due to complaints regarding Tibal Impactor pad fracture.

Firm Action

On 10/14/2014, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected distributors and hospital Risk Managers. All distributors were notified via electronic mail, and all hospital Risk Managers and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call the customer call center at 1-877-946-2761.

Root Cause

Labeling design

Distribution Pattern

Worldwide Distribution - US Nationwide in the state of AL, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, ND, NH, NJ, NV, NY, OH, OR, PA, SC, TX, VA, WI, NC, TN, OK, AZ, UT, AK, PR, DC. and the countries of: Germany, China, Canada, Australia, Japan, United Arab Emirates, Austria, Belgium, Switzerland, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Iceland, Italy, Malta, Netherlands, Norway, Poland, Russian Fed., Saudi Arabia, Sweden, Slovakia, and South Africa.