OviTex 1S Reinforced BioScaffold 10x12cm, Part Number F10256-1012G
FDA Recall Z-0423-2019 · Posted None
Advertisement
Recalling FirmAROA Biosurgery, LTD.
Recall StatusTerminated
Product CodeFTM
Date Initiated2018-04-06
Date Terminated2019-04-15
Product Quantity881 total
Reason for Recall
Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
Firm Action
Beginning in April 2018, customers were visited by representatives and the affected units were replaced.
Root Cause
Other
Distribution Pattern
The products were distributed to the following US states: AL, CA, FL, IN, MA, MI, NH, and NY.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.