Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic Nuu N5004L, Horizon PDM, US PT-000429: Cable, Charging, Omnipod 5, Horizon PDM

FDA Recall Z-0423-2023 · Posted 2022-12-07

Recalling FirmInsulet Corporation
Recall StatusOpen, Classified
Product CodeQFG
Date Initiated2022-11-14
Product Quantity129,086

Reason for Recall

The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.

Firm Action

On November 14, 2022, the firm initiated their Press Release at https://investor.insulet.com/news-releases/news-release-details/insulet-issues-nationwide-voluntary-medical-device-correction and notified affected customers through Urgent Medical Device Correction letters. Customers were instructed to monitor their Omnipod 5 Controller for the following issues: - Charging port or charging cable appears melted, deformed, or discolored - Controller experiences extreme overheating (uncomfortable to hold) or emits an odor while charging If customers experience any of the issues above, they should disconnect the charger from the power outlet and do NOT charge the Omnipod 5 Controller. Promptly contact Insulet s dedicated Customer Care team specifically trained to help you at 1-800-641-2049 to request a replacement device. If you have a compatible Android smartphone (please see omnipod.com/compatibility), you may want to switch to using the Omnipod 5 App on your smartphone. Customers were provided with additional precautions to take to continue use of the device.

Root Cause

Device Design

Distribution Pattern

Domestic distribution nationwide.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑