HF Resection Electrodes, Model Numbers WA22602S, WA22603S, WA22621S and WA22657S

FDA Recall Z-0424-2023 · Posted 2022-12-07

Recall StatusOpen, Classified
Product CodeFJL
Date Initiated2022-11-04
Product Quantity186 boxes (930 pieces)

Reason for Recall

An incompatible HF cable may be packaged with the HF Resection Electrodes. This cable cannot be connected to the electrosurgical generator and may lead to delay or cancellation of the surgical procedure.

Firm Action

On November 4, 2022, the firm notified affected customers. Customers were instructed to identify any affected product in their inventory. Affected product should be returned to Olympus, who will issue a credit to the customer upon return of affected product. Those who have distributed the devices outside their facility should notify their customers of this removal action immediately. To return product, contact your Olympus customer service representative at 1-800-848-9024.

Root Cause

Packaging process control

Distribution Pattern

US Nationwide domestic distribution.