COOK Stiffened Micropuncture Introducer Set, 4 FR. Product Reorder Number: MPIS-401-10.0-SC
FDA Recall Z-0425-03 · Posted 2003-02-28
Advertisement
Recalling FirmCook, Inc.
Recall StatusTerminated
Product CodeDYB
Date Initiated2002-11-25
Date Terminated2003-03-21
Reason for Recall
The tip of the stiffened cannula might break during use
and remain in the patient.
Firm Action
Recall letters were sent to each customer on November 26, 2002 advising them of the recall and to return stocks on hand.
Root Cause
Other
Distribution Pattern
United States, Brazil, Canada, Dominican Republic and Israel.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.