Terminated

Signa Ovation, 0.35T MR Systems (MFO 1-4). Indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the musculoskeletal, vascular, cardiac, and neuro systems.

FDA Recall Z-0426-2011 · Posted 2010-11-19

Recalling FirmGE Healthcare, LLC
Recall StatusTerminated
Product CodeLNH
Date Initiated2010-08-23
Date Terminated2012-07-13
Product Quantity202

Reason for Recall

If the lateral table adjustment crank handle breaks off, it has sufficient attractive force to the magnet and may result in patient injury.

Firm Action

Consignees were sent on 8/23/10 a GE "Urgent Medical Device Correction" letter dated August 20, 2010. The letter was addressed to Hospital Administrators / Risk Mangers, Radiology Department Managers and Radiologists. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product Correction, and Contact Information. Customers were asked to discontinue use of their MR system and to contact their local service representative, if the lateral table adjustment crank handle on the table becomes loose.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution -- USA, including states of AL, AZ, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, LA, KS, KY, LA, MD, MA, MI, MN, MS,MO,MT, NV, NH, NJ, NY, NC, OH, OK, OR, PA, PR, SC, SD, TN, TX, VA, WA, WV, and WI and countries of VENEZUELA, ECUADOR, UNITED KINGDOM, UNITED ARAB EMIRATES, TURKEY, SPAIN, RUSSIAN FEDERATION, REPUBLIC OF KOREA, POLAND, PHILIPPINE,S OMAN, MEXICO, KUWAIT, JAPAN, ITALY, HUNGARY, GERMANY, EGYPT, DENMARK, CHILE, CHINA, and BRASIL.