Cutera brand truSculpt infrared System (optional), Massager, Vacuum, radio frequency induced heat, Product is manufactured and distributed by Cutera, Inc., Brisbane, CA The truSculpt massage device is intended to provide a temporary reduction in the appearance of cellulite. The truSculpt infrared (optional) and RF energies are intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation.
FDA Recall Z-0427-2014 · Posted 2013-12-09
Advertisement
Recalling FirmCutera, Inc.
Recall StatusTerminated
Product CodeOPT
Date Initiated2013-10-16
Date Terminated2013-12-09
Product Quantity451
Reason for Recall
Cutera received two adverse events associated with the edge of the electrode, when it was pressed against neck skin over the thyroid cartilage and due to the absence of adequate underlying conductive tissue, increased skin heating occurred. Both patients developed linear welts and blisters consistent with partial thickness thermal burns to the skin.
Firm Action
Cutera sent an IMPORTANT SAFETY INFORMATION letter dated October 16, 2013, to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.
It is important that you share this correspondence and the updated Operator Manual andvtreatment guidelines with all staff at your office that use the truSculpt, so they are aware of this new safety information. Please immediately remove and dispose of the current truSculpt Operator Manual and clinical guidelines and replace them with the new and revised truSculpt Operator Manual and clinical guidelines.
Customers with questions were instructed to call 415-657-5500 or 415-657-5568.
Root Cause
Employee error
Distribution Pattern
Worldwide Distribution - USA (nationwide) and Internationally to Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.