Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Depuy Preservation Uni Knee All Poly Tibia, RM/LL, Sz 4, 11.5 mm, sterile; Product Code/REF 1498-11-410.

FDA Recall Z-0429-2008 · Posted 2007-12-06

Recall StatusTerminated
Product CodeNJL
Date Initiated2007-09-27
Date Terminated2008-04-30
Product Quantity9

Reason for Recall

An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.

Firm Action

Consignee hospitals and implanting physicians were notified of the recall via an Urgent Information-Recall Notice letter and Dear Doctor letters dated 9/27/07. The hospitals were instructed to determine whether they had any of the recalled products and to contact their sales representative to pick up the product. Depuy sales representatives were notified to collect the products from consignee hospitals via email on 10/1/07. The decision to discuss this product defect with patients who had already been implanted was left to the implanting physicians.

Root Cause

Packaging process control

Distribution Pattern

Nationwide.