BD PhoenixSpec Calibrator Kit, consisting of 3 tubes identified as 0.25, 1.0 and 4.0 McFarland unit calibrators, and labeled in part ***Becton, Dickinson and Company, 7 Loveton Circle, Sparks, MD 21152 USA***

FDA Recall Z-0430-06 · Posted 2006-01-24

Recall StatusTerminated
Product CodeJQX
Date Initiated2005-12-01
Date Terminated2006-04-06
Product Quantity145 kits

Reason for Recall

A calibrator for an in vitro diagnostic kit is not standarized correctly and may cause a calibration error when measuring microbiological cultures for turbidity.

Firm Action

The recalling firm notified consignees by phone, fax and e-mail on 12/01/05. The notification advised that nephelometers displaying 'CAL' error messages on the LED should not be used until calibrated with a different kit lot. Consignees were requested to discard referenced lots for replacement. A response form was included in the notification.

Root Cause

Other

Distribution Pattern

Product was sold to direct accounts nationwide and international affiliated distributors in Japan, Belize, Taiwan, Canada, Korea, Australia, and the Philippines.