Terminated

Dimension Vista(R) 500 and Dimension Vista(R) 1500 - Reloaded Vista Vials - limited to systems using software versions 3.4 and 3.5.1. Diagnostic device intended to determine a variety of analytes in human body fluids.

FDA Recall Z-0430-2014 · Posted 2013-12-02

Recall StatusTerminated
Product CodeJJE
Date Initiated2013-05-20
Date Terminated2014-04-10
Product Quantity776

Reason for Recall

There is the potential for the vial volume of "in use" Vista calibrators, controls, or sample diluent to incorrectly revert to full volume when removed and reloaded. Dimension Vista(R) 500 and Dimension Vista(R) 1500 - Reloaded Vista Vials - limited to systems using software versions 3.4 and 3.5.1

Firm Action

A notification letter was sent to consignees in May 2013.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution: USA (nationwide) and iinternationally to Canada.