Terminated

Sex Hormone Binding Globulin (SHBG) Reagent for use on Access Immunoassay Systems, Part number(s): A48617 The Access SHBG assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of Sex Hormone Binding Globulin levels in human serum and plasma using the Access Immunoassay Systems.

FDA Recall Z-0431-2011 · Posted 2010-11-22

Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeCDZ
Date Initiated2010-07-26
Date Terminated2012-05-04
Product Quantity1,113 Units ( 894 in US & 219 in Canada)

Reason for Recall

The recall was initiated because the Instruction for Use (IFU) for Access SHBG does not specify the units of measure to be used when calculating the Free Androgen Index (FAI %). Access SHBG (nmol/L) values and Access Testosterone (ng/mL) values are reported in different default units. The testosterone value (when reported in ng/mL) requires conversion to nmol/L before the FAI % can be calculated.

Firm Action

Beckman Coulter sent an Urgent Product Corrective Action (PCA) letter on the week of July 26, 2010 to the affected customers. The letter was dated July 30, 2010, and identified the product, the problem and provided the appropriate equation to recalculate the FAI% values. The customers were asked to share this information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. Customers were also asked to complete and return the enclosed response form within 10 days to confirm that they had received the important notification. The customers were instructed to contact Beckman Coulter Technical Support at 1-800-854-3633 in the United States, or contact their local Beckman Coulter Representative if they had any questions regarding the notification.

Root Cause

Pending

Distribution Pattern