Harvard 1 Single Channel Syringe Infusion Pump, P/N 2003-001
FDA Recall Z-0432-05 · Posted 2005-01-26
Advertisement
Recalling FirmHarvard Clinical Technology
Recall StatusTerminated
Product CodeMEA
Date Initiated2004-11-03
Date Terminated2012-05-07
Product Quantity154 units
Reason for Recall
Potential pump motor/encoder assembly failure may cause over infusing medication in patients
Firm Action
Harvard Clinical Technology notified consignees by letter on 11/3/04. Harvard will contact consignees by telephone to schedule a return. The motors inside the recalled products will be replaced at Harvard Clinical. Harvard will provide Loaner pumps upon request during the repair process.
Root Cause
Other
Distribution Pattern
AK, KY, TX, MA, NY, CA, GA, OK, IL, CT, LA, FL, MD,NC, NJ, VA, MI, WI, IA, WV,OH,