Brand Name: Klassic BiPolar Head Model/Catalog Number: 3205.41.022 Klassic BiPolar Head, 41 mm OD, 22 mm Head; 3205.42.022 Klassic BiPolar Head, 42 mm OD, 22 mm Head; 3205.43.022 Klassic BiPolar Head, 43 mm OD, 22 mm Head; 3205.44.028 Klassic BiPolar Head, 44 mm OD, 28 mm Head; 3205.45.028 Klassic BiPolar Head, 45 mm OD, 28 mm Head; 3205.46.028 Klassic BiPolar Head, 46 mm OD, 28 mm Head; 3205.47.028 Klassic BiPolar Head, 47 mm OD, 28 mm Head; 3205.48.028 Klassic BiPolar Head, 48 mm OD, 28 mm Head; 3205.49.028 Klassic BiPolar Head, 49 mm OD, 28 mm Head; 3205.50.028 Klassic BiPolar Head, 50 mm OD, 28 mm Head; 3205.51.028 Klassic BiPolar Head, 51 mm OD, 28 mm Head; 3205.52.028 Klassic BiPolar Head, 52 mm OD, 28 mm Head; 3205.53.028 Klassic BiPolar Head, 53 mm OD, 28 mm Head; 3205.54.028 Klassic BiPolar Head, 54 mm OD, 28 mm Head; 3205.55.028 Klassic BiPolar Head, 55 mm OD, 28 mm Head; 3205.56.028 Klassic BiPolar Head, 56 mm OD, 28 mm Head; 3205.58.028 Klassic BiPolar Head, 58 mm OD, 28 mm Head; 3205.60.028 Klassic BiPolar Head, 60 mm OD, 28 mm Head
FDA Recall Z-0432-2026 · Posted 2025-11-05
Advertisement
Recalling FirmTotal Joint Orthopedics, Inc.
Recall StatusOpen, Classified
Product CodeKWY
Date Initiated2025-09-23
Product Quantity103 units
Reason for Recall
Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.
Firm Action
On September 5, 2025 Total Joint issued customer a Field action notification. On September 23, 2025, Total Joint issued a Urgent Medical Device Recall Notification. Total Joint ask consignees to take the following actions:
1. Stop implanting the affected product.
2. Locate and quarantine all affected inventory.
3. With preliminary info we initially sent a Market Withdrawal notification on 9/05/2025 to initiate the quarantining and return of all BiPolar Heads in your possession.
4. A list of all BiPolar Heads in your possession as of 9/05/2025 is shown within the letter, if you have already returned them we appreciate your swift response, if not please return these items immediately.
5. Please complete and return the attached acknowledgment form confirming receipt of this notice and detailing the quantities returned. If inventory discrepancies are identified, please note that in the Return Response Box.
6. Please contact Total Joint Orthopedics at (801)486-6070 with any questions or concerns.
Root Cause
Packaging
Distribution Pattern
US Nationwide distribution in the states of Ohio, Texas, Florida, Georgia, and North Carolina.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.