Mach 1 6F JR 3.5 Guide Catheter. Catalog No. 34356-685. Sterilized with ethylene oxide gas. Nonpyrogenic. Intended for one procedure use only. Manufacture: Boston Scientific Scimed, Inc., One Scimed Place, Maple Grove, MN USA 55311
FDA Recall Z-0433-06 · Posted 2006-01-24
Recalling FirmBoston Scientific Scimed
Recall StatusTerminated
Product CodeDQY
Date Initiated2004-05-24
Date Terminated2007-01-23
Product Quantity10 devices
Reason for Recall
Some Mach 1 6F Coronary Guide Catheter could be mislabeled in that a few units labeled as Mach 1 6F Kimny Curve style, could contain a Mach 1 6F JR 3.5 curve style and a few units labeled as Mach 1 6F JR 3.5 curve style, could contain a Mach 1 6F Kimny curve style.
Firm Action
A Recall Notice, dated 05/28/05, was sent to Risk Managers. The notice informed them of the issue and instructed them to discontinue use and segregate affected units. A Tracking/verification Form is requested to be returned. Arrangements will be made for returning affected product.
Root Cause
Other
Distribution Pattern
Within the US to GA, MS, NC ,TN. OUS to include: Canada, Hong Kong, Italy, Netherlands, Saudi Arabia, United Kingdom