Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the United States).

FDA Recall Z-0433-2007 · Posted 2007-02-14

Recall StatusTerminated
Product CodeDTQ
Date Initiated2006-10-16
Date Terminated2009-07-07
Product Quantity207

Reason for Recall

A 'battery needs service' message may be displayed, or the remaining battery time may incorrectly become zero, even when the battery is fully functional.

Firm Action

Consignees were sent a letter dated 11/10/06 informing them of the nature of the issues and that the firm would contact them to schedule a service call to install the software upgrade package. The software upgrade installation began at customers on 10/16/06, prior to issuance of the letter to consignees.

Root Cause

Other

Distribution Pattern

Worldwide distribution ---- including USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.