VISTA BRITE TIP RDC(1) Guiding Catheter. Product Code 67021055, lot X0704002, Precutaneous Catheter.
FDA Recall Z-0434-05 · Posted 2005-01-26
Advertisement
Recalling FirmCordis Corporation
Recall StatusTerminated
Product CodeDQY
Date Initiated2004-12-07
Date Terminated2008-11-19
Product Quantity153 units
Reason for Recall
Two complaints have been received and confirmed, regarding a lot, indicating that the distal end had an 'RDC' shape rather than the 'RDC(1)' shape. The RDC shape has a larger curve. The product was manufactured with the incorrect forming wire during the guiding catheter manufacturing process.
Firm Action
Packets containing the Customer Letter and Acknowledgement Form for each domestic consignee were sent by overnight delivery to the appropriate Cordis Endovascular sales representatives. The letter is being hand-delivered to each initial domestic consignee by the Cordis Endovascular sales representative. Any remaining units at the account will be removed from the shelf and returned to Cordis. The account will be asked to sign the Acknowledgement Form.
Root Cause
Other
Distribution Pattern
Distributed to 57 accounts throughout the United States and three international accounts.