IO FiX X-Post Reamer, Extremity Medical, www.extremitymedical.com; Customer Service: 888.499.0079 The X-Post Reamer is an accessory to the IO FiX Screw and Washer System that is used for reaming bone most common to orthopedic surgical procedures.

FDA Recall Z-0438-2013 · Posted 2012-11-26

Recalling FirmExtremity Medical LLC
Recall StatusTerminated
Product CodeHTO
Date Initiated2011-04-25
Date Terminated2013-03-21
Product Quantity156 units

Reason for Recall

Product Event Reports revealed a broken reamer experienced intra-operatively. Pre-drilling was not performed prior to reaming.

Firm Action

Extremity Medical sent a notification by e-mail that describes the "Slight modification to the Surgical Technique" on April 25, 2011. For questions regarading this recall call 973-588-8980.

Root Cause

Employee error

Distribution Pattern

Worldwide Distribution - USA including FL, MA, NC, CA, MD, VA, IN, AL, NJ, AZ, OK, MT, NY, OH, IA, MI, TX, WI, GA, IL, TN and NE. Internationally to Switzerland.