Open, Classified

CardioQuip Modular Cooler-Heater, Model:MCH-1000(m), Power: 115V-60hz,16A

FDA Recall Z-0438-2024 · Posted 2023-12-01

Recalling FirmCardioQuip, LLC
Recall StatusOpen, Classified
Product CodeDWC
Date Initiated2023-10-19
Product Quantity150 units

Reason for Recall

There is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components.

Firm Action

On 10/19/23, CardioQuip issued a "Urgent: Medical Device Correction" notification to all affected consignees. This notification was an update to the firm's May 22, 2023 communication. CardioQuip ask consignees to take the following actions: 1. Prime the MCH-1000(i) and MCH-1000(m) prior to each use. 2.Never operate the device dry (see communication for further details). 3. Design Changes to be implemented (see communication for further details). A. Relocation of the water Level Sensor (see communication for further details). B. Software Update (see communication for further details). 4. Please fill out and submit the receipt of understanding of the contents of this letter to CardioQuip within 5 business days or receipt. 5. Any adverse events experienced relating to CardioQuip devices should be reported to CardioQuip or through the FDA s MedWatch Program: Phone: +1 (800) FDA-1088 Web: www.fda.gov/medwatch/report.htm

Root Cause

Device Design

Distribution Pattern

US Distribution: TX, IN, IL, WI, FL, NJ, PA, KY, VA, WV, CT, CA, WA, AL, IA

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →