CardioQuip Modular Cooler-Heater, Model:MCH-1000(i), Power: 115V-60hz,16A
FDA Recall Z-0439-2024 · Posted 2023-12-01
Advertisement
Recalling FirmCardioQuip, LLC
Recall StatusOpen, Classified
Product CodeDWC
Date Initiated2023-10-19
Product Quantity125 units
Reason for Recall
There is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components.
Firm Action
On 10/19/23, CardioQuip issued a "Urgent: Medical Device Correction" notification to all affected consignees. This notification was an update to the firm's May 22, 2023 communication. CardioQuip ask consignees to take the following actions:
1. Prime the MCH-1000(i) and MCH-1000(m) prior to each use.
2.Never operate the device dry (see communication for further details).
3. Design Changes to be implemented (see communication for further details).
A. Relocation of the water Level Sensor (see communication for further details).
B. Software Update (see communication for further details).
4. Please fill out and submit the receipt of understanding of the contents of this letter to CardioQuip within 5 business days or receipt.
5. Any adverse events experienced relating to CardioQuip devices should be reported to CardioQuip or through the FDA s MedWatch Program: Phone: +1 (800) FDA-1088 Web: www.fda.gov/medwatch/report.htm
Root Cause
Device Design
Distribution Pattern
US Distribution: TX, IN, IL, WI, FL, NJ, PA, KY, VA, WV, CT, CA, WA, AL, IA
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.