Mammotome EX Holster

FDA Recall Z-0440-05 · Posted 2005-01-27

Recalling FirmEthicon Endo-Surgery
Recall StatusTerminated
Product CodeKNW
Date Initiated2004-12-10
Date Terminated2008-10-29
Product Quantity120 devices

Reason for Recall

A capacitor of the wrong value was used. The capacitor used was 0.11 microfarad instead of 0.22 microfarad called for in the product specifications.

Firm Action

The recalling firm notified hospitals and distributors of the product by letter, dated 12/10/04.

Root Cause

Other

Distribution Pattern

The product was distributed to consignees located in MI, TN, IL, NC, GA, IL, CA, FL, IN, WI, TX, AR, KS, IA, OK, NY, PA, MS, KY, ME, MO, MA, VA, NJ, LA, MN, SD, UT, CO, AL, and NH. There are no foreign consignees.