8MM, LARGE HEM-O-LOK CLIP APPLIER; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

FDA Recall Z-0441-2015 · Posted 2014-11-25

Recall StatusTerminated
Product CodeNAY
Date Initiated2014-11-03
Date Terminated2015-02-03
Product Quantity509 total units, all affected devices

Firm Action

Urgent Medical Device Recall letters were sent to affected sites on November 3, 2014. Letters inform customers of the reason for recall, risk to health and the affected instrument. A table of all of the affected IS4000 instruments affected is attached to the letter indicating the part number and product name for each. Letters also inform customers that all instruments shipped since July 2014 are manufactured with a new material and not subject to recall. Customers are to take the following actions: 1. Ensure that all affected personnel are aware of the information. 2. Identify affected product in their possession 3. Contact ISI rep for replacement. 4. Complete and return the attached Acknowledgement From. Until replacement instruments are received, the old instruments may be used provided they do not show signs of degradation and that the reprocessing instructions are followed. Questions to 800-876-1310 Option 3 ( 6 am to 5 pm PST) or [email protected].

Root Cause

Labeling design

Distribution Pattern

US Distribution -- FL, TX, NY, NV, OR, IN, UT, and MA.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑