ELEKTA LIMITED Elekta Synergy, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
FDA Recall Z-0443-2024 · Posted 2023-12-01
Advertisement
Recalling FirmElekta, Inc.
Recall StatusOpen, Classified
Product CodeIYE
Date Initiated2023-10-25
Product Quantity1019 units
Reason for Recall
Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
Firm Action
Elekta released Field Safety Notice to its consignees on the 10/25/2023 via email. The notice explained the issue, potential risk, and provides information for the safe use of the instrument until Elekta Field Service Representatives can perform the correction.
Root Cause
Under Investigation by firm
Distribution Pattern
worldwide