REMstar Heated Humidifer Main PCA Repair Kit, Part number 1006265

FDA Recall Z-0447-2007 · Posted 2007-01-10

Recalling FirmRespironics, Inc.
Recall StatusTerminated
Product CodeBTT
Date Initiated2006-10-25
Date Terminated2009-12-15
Product Quantity235 units

Reason for Recall

AC inlet connector failure that can result in thermal damage to the humidifier enclosure and/or ignition of materials external to the device

Firm Action

The recalling firm issued a recall letter dated 10/25/06 to their domestic and Canadian dealers informing them of the problem and the need to contact the end users. These dealers were given two options: 1) contact the end user by forwarding a letter the recalling firm provided or 2) contacting the third party company with their distribution list. The recalling firm issued a recall letter dated 10/25/06 regarding the service repair kit and the need to contact end users that had the device repaired using this kit. The recalling firm issued a recall letter date 10/25/06 to the international customers informing them of the problem, contact the end users, and the need to destroy the recalled part.

Root Cause

Other

Distribution Pattern

Worldwide Distribution ---- including USA and countries of Denmark, Austria, France, Canada, Netherlands, Japan, Hong Kong, Austrailia, Great Britain, Spain, Belgium, Germany, Poland, Ireland, Portugal, Saudia Arabia, Korea, Mexico, Finland, Taiwan, Philippines, Iceland, Brazil, Kuwait, New Zealand, Chile, India, Egypt, Norway, Turkey, Jordan, Thailand, Hungary, Russia, Peru, South Africia, Israel, Czech, Pakistan, Lebanon, Venezuela, Sweden, United Arab Emirates, Qatar, Slovakia, Bulgaria, Bermuda, and Colombia.