Remel TB Potassium Permanganate (.5% Aqueous) 40192, packaged in 250 ml bottles, 5/pkg. A stain for use in qualitative procedures as a counterstain in the fluorescent microscopic detection of mycobacteria.

FDA Recall Z-0449-2015 · Posted 2014-11-25

Recalling FirmRemel Inc
Recall StatusTerminated
Product CodeJTS
Date Initiated2014-10-29
Date Terminated2015-06-01
Product Quantity79/5/250- packages/bottles/ml

Reason for Recall

The product may appear cloudy or discolored and may not perform correctly in presumptive diagnosis of mycobacterial disease.

Firm Action

Thermo Fisher Scientific sent a Medical Device Recall letter dated October 29, 2014, to all affected customers. The letter notified customers that Remel, Inc., a part of Thermo Fisher Scientific, was voluntarily recalling the affected product. Customers were instructed to notify any personnel who need to be notified of the potential for false negative results. Customers should review results reported using this lot and consider retesting and/or seek appropriate expert advice. Return the attached Medical Device Recall Return Response form acknowledging receipt of the notice and disposal of the affected product. Customers with questions were instructed to contact the Technical Services Department at 800-255-6730 (US) or 913-888-0939 (International). For questions regarding this recall call 913-888-0939.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution to USA including AL, CA, CO, IA, MO, NC, NY, NV, OH, PA, SC, SD, TN, TX, VA, WI and Internationally to Canada.