Freedom Boom models series: F110, F120, F130, F320, F330, F350, F410, and F420 with built in 42A Lightweight Spring Arm part numbers 42A-43L, 42A-43M, 42A-43H, and 42A-43S. Ceiling mounted utilities and equipment dispensing system.

FDA Recall Z-0452-2024 · Posted 2023-12-01

Recalling FirmIHB OPERATIONS B.V.
Recall StatusOpen, Classified
Product CodeFQO
Date Initiated2023-07-19
Product Quantity1,497

Reason for Recall

Weakened connection point on the adjustable suspension arm, potentially causing damage or a break.

Firm Action

A recall notification issued October 09, 2023 via certified mail . Follow-up will be made by service representatives within 2 weeks to arrange for correction to replace all bearing assemblies in the field with an updated design. Prior to the field correction, inspect your Freedom 42A Lightweight Spring Arm for any movement in the end bearing assembly. If any loose connections are identified, please immediately remove the Spring Arm from use and notify your local authorized representative. Address questions or concerns to your local authorized representative or Skytron at 1-800-759-8766.

Root Cause

Device Design

Distribution Pattern

US: AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MB, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, VT, WA, WI, WV.