Cepheid brand Xpert MRSA/SA-BC US-IVD, an In-vitro diagnostic test kit, Catalog Numbers: GXMRSA/SA-BC-10, Lots Affected: 2803 and 3001;
FDA Recall Z-0453-2012 · Posted 2012-01-11
Advertisement
Recalling FirmCepheid
Recall StatusTerminated
Product CodeNQX
Date Initiated2011-11-17
Date Terminated2012-01-09
Product Quantity1289
Reason for Recall
Cepheid has received reports from some of its customers regarding higher than desired (>5%) rates of invalid results when using lot numbers 2803 and 3001 of the Xpert MRSA/SA BC (blood culture) product.
Firm Action
On 11/17/11, Cepheid issued letters via mail to all its direct consignees, informing them of the affected products and providing instructions on the recall.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Product was released for distribution in US.