Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Record updated 3/19/26 to include: CARESCAPE B450 MBA313 2095800-001 CARESCAPE MONITOR B450 (ATO MODEL) 2068491-002 CARESCAPE MONITOR B450 B1 2107633-001 CARESCAPE MONITOR B450-GS (V2.0.8) 2094080-001
FDA Recall Z-0453-2026 · Posted 2025-11-07
Advertisement
Recalling FirmGE Healthcare Finland Oy
Recall StatusOpen, Classified
Product CodeMHX
Date Initiated2025-09-26
Product Quantity3271
Reason for Recall
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
Firm Action
On September 26, 2025, the recalling firm began notifying affected customers. A second letter was issued on October 3, 2025 which addressed a correction to Affected Product Details information.
The customer letter advised users to use a monitor with a battery that is not within the affected battery population. However, pending correction by GE HealthCare, users may continue to use a monitor with affected batteries by following instructions included in the letter. These instructions include having mains power connected whenever possible, confirming appropriate battery status prior to use, and being aware that the monitor may still shut down during patient transport. Users are advised to be prepared to provide alternate methods of assessing the clinical status of the patient, bring a mains power cord during patient transport, and bring a spare battery.
GE HealthCare will correct all affected products at no cost to the customer. A GE HealthCare representative will contact each customer to arrange for the correction.
If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Root Cause
Component change control
Distribution Pattern
Worldwide - US Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.