ThermoSuit Manufactured for Life Recovery Systems. HD, LLC, 170 Kinnelon Rd. Kinnelon, NJ 07405 USA Tel.: (973) 283-2800 FAx.: (973) 283-2910 Temperature reduction in patients where clinically indicated; Monitoring of patient temperature
FDA Recall Z-0456-2012 · Posted 2012-01-11
Advertisement
Recalling FirmLife Recovery Systems HD, LLC
Recall StatusTerminated
Product CodeNZE
Date Initiated2011-10-21
Date Terminated2013-01-24
Product Quantity28 units
Reason for Recall
ThermoSuit System Model TSP-50 Pump/Controller may have corrosion on the temperature circuit board which may result in incorrect patient temperature reading.
Firm Action
Life Recovery Systems sent recall letters/return response forms dated 10/24/2011 to all consignees. Life Recovery Systems has expanded their recall to include additional units and to provide additional service to products out in the field. A new recall letter was sent to consignees on 4/27/2012.
Root Cause
Process control
Distribution Pattern
Nationwide Distribution including NY, NC, LA, NJ, AL. PA TX, KY, MO, MA, and TN