DA-2006P - Defibrillator / Pacer Analyzer; Product Usage: Intended to measure the output of a defibrillator or transcutaneous pacemaker.

FDA Recall Z-0457-2016 · Posted 2015-12-22

Recall StatusTerminated
Product CodeDRL
Date Initiated2015-10-28
Date Terminated2017-02-15
Product Quantity5 units

Reason for Recall

Device is not functioning as intended: Two wires running to the ECG distribution circuit board are reversed resulting in inverted waveforms.

Firm Action

The recalling firm issued return authorization numbers to the two customers who submitted complaints regarding the malfunctioning device. The customers sent back the units and the recalling firm replaced their devices with new units. On 10/29/2015, the recalling firm contacted their third customer by phone and informed them that the three devices they had in their possession were miswired which would cause the inverted ECG waveform. The recalling firm issued a return authorization number to their customer for the three devices. The recalling firm told their customer they would repair the devices and return them.

Root Cause

Employee error

Distribution Pattern

US Nationwide Distribution in the states of South Carolina, Florida, and Canada.