Recall StatusOpen, Classified
Date Initiated2021-11-19
Product Quantity29495 kits
Reason for Recall
A subset of affected discs are at risk of leakage when it is used more than one time with Simplexa COVID-19 Direct and/or Simplexa Flu A/B & RSV Direct Gen II. Leakage from a well that had a positive sample may cause further cross-contamination in areas of the lab if not properly cleaned.
Firm Action
On 11/19/2021, the firm notified affected customers via email of "URGENT: MEDICAL DEVICE CORRECTION NOTIFICATION." The letter indicated, "Actions to be taken by the Customer/User/Distributor:
- Any remaining discs in inventory that are listed in the table below should only be used one time and not reused.
- Any sample with an error code should be retested.
- If leakage is suspected, the spillage should be removed and the instrument surface should be decontaminated. Ensure the practice of Good Laboratory Procedures including regular cleaning.
- Notify others within your network that may have received these kits.
- Complete and return the Acknowledgement and Receipt Form (page 3) of this notification letter.
- If your firm has further distributed the devices, please share a copy of this correction notification and response form with those customers and provide a copy of the completed response form back to us.
If you have any questions or if you would like replacement product for the Direct Amplification Discs you still have in stock, please contact DiaSorin Molecular Technical Service, Monday Friday, 7:00am to 5:00pm (Pacific Time) at (800) 838-4548, option 3, or by email at
[email protected]."
Root Cause
Under Investigation by firm
Distribution Pattern
US:
AZ,
CA,
CO,
CT,
DC,
FL,
GA,
HI,
IA,
IL,
IN,
KS,
KY,
LA,
MA,
MD,
ME,
MI,
MN,
MO,
NC,
ND,
NE,
NH,
NJ,
NM,
NV,
NY,
OH,
PA,
RI,
SC,
SD,
TN,
TX,
VA,
VT,
WA,
WI,
WV; and
Worldwide:
Kuwait,
Australia,
Israel,
Italy,
Canada,
Chile,
Puerto Rico,
United Arab Emirates.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.