QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-10X3-E, Order Number G22667; used to dilate strictures of the biliary tree
FDA Recall Z-0458-2024 · Posted 2023-12-01
Advertisement
Recalling FirmWilson-Cook Medical Inc.
Recall StatusOpen, Classified
Product CodeFGE
Date Initiated2023-11-03
Product Quantity111 devices
Reason for Recall
Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally.
Firm Action
Cook Medical issued an URGENT: MEDICAL DEVICE REMOVAL notice on 11/03/2023 by mail. The notice explained the problem with the device, risk to health, and requested the following actions be taken:
Customer: Cease use, quarantine, and return all affected devices. If devices were further distributed, those to whom the devices were sent must be notified.
Distributor: Quarantine all affected devices, notify your customers to whom you distributed devices. Return all devices to Cook Medical.
Root Cause
Under Investigation by firm
Distribution Pattern
US
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.