The Dimension TSHL Flex reagent cartridge (Thyroid Stimulating Hormone) manufactured by Siemens Healthcare Diagnostics, Inc., Newark, DE.
FDA Recall Z-0459-2011 · Posted 2010-11-24
Advertisement
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Recall StatusTerminated
Product CodeJLW
Date Initiated2010-04-12
Date Terminated2013-03-19
Product Quantity4187
Reason for Recall
Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results.
Firm Action
Siemens issued an "Urgent Field Safety Notice" dated April 2010, informing users of the problem and instructing them to inspect cartridges for leaking fluid. Cartridges with leaking fluid should be reported to the Siemens Technical Solutions Center. The Technical Solutions Center can be contacted at 800 441 9250.
Root Cause
Other
Distribution Pattern
Worldwide Distribution: USA, Europe, Australia, China, Japan, Korea, and Saudi Arabia.